Cerabar T
10 Endress+Hauser
Operating conditions (environment)
Ambient temperature range PMC131
–20 to +85 °C
PMP131 and PMP135
For non-hazardous areas: –25 to +70 °C
Ex i: –25 to +65 °C
Storage temperature range PMC131
–50 to +100 °C
PMP131 and PMP135
–40 to +85 °C
Climate class PMC131
4K4H as per DIN EN 60721-3
PMP131 and PMP135
4Z
with Z = 70 °C as per VDI/VDE 3540
Degree of protection PMC131
Plug M 16 x 1,5 (DIN 43650/A), ½ NPT: IP 65/NEMA 4X
Plug M12x1: IP 65/ NEMA 4
Cable: IP 68/NEMA 6P (1 mWS/24 h)
PMP131 and PMP135
Plug M 16 x 1,5 (DIN 43650/A), ½ NPT: IP 65/NEMA 4X
Plug M 12x1 and gauge pressure sensors: IP 65/NEMA 4X
Plug M 12x1 and absolute pressure sensors: IP 68/NEMA 6P (1 mWS/24 h)
Cable: IP 68/NEMA 6P (1 mWS/24 h)
Vibration resistance 4M5 as per DIN EN 60721-3
Electromagnetic compatibility PMC131
Interference emission as per EN 61326 electrical device class B,
Interference immunity as per EN 61326 appendix A (industrial use).
PMP131 and PMP135
Electromagnetic combatibility according to EN 61326;
For details refer to the declaration of conformity (see ATEX Safety Instructions XA142P)