Cerabar T
Endress+Hauser 5
Power max. 6 W
Switch frequency max. 10 Hz
Input PLC • Input resistance R
i
≤ 2 kΩ
• Input current I
i
≥ 10 mA
Inductive loads To prevent electrical interference, only operate an inductive load (relays, contactors, solenoid valves) when
directly connected to a protective circuit (free-wheeling diode or capacitor).
Power supply
PMC131 Plug/cable connection
P01-PMC131xx-04-xx-xx-xx-001
Electrical connection: Analog/current output
P01-PMC131xx-04-xx-xx-xx-002
1 Plug M 16 x 1.5 (DIN 43650/A), ½ NPT
2 Plug M 12 x 1
3 Cable (rd = red, bk = black, gnye = green-yellow)
Plug M 16 x 1.5 (DIN 43650/A), ½ NPT Plug M 12x1 5 m/25 m cable
➀ Plug-in housing ➀ Reference pressure line
➁ Gasket
M16 / ½ NPT
M12x1
5 m/25 m
1
2
ENDRESS+HAUSER
CERABAR T
ENDRESS+HAUSER
CERABAR T
ENDRESS+HAUSER
CERABAR T
1
–
–
–
+
+
+
+
PE
–
rd
bk
gnye
2
1
1+
2–
12...30 V
12...30 V
12...30 V
4...20 mA
4...20 mA
4...20 mA
1
2
3
1+
2
3-