Cerabar T
6 Endress+Hauser
PMP131 and PMP135 Plug/cable connection
P01-PMP13xxx-04-xx-xx-xx-001
Electrical connection: Analog/current output
P01-PMP13xxx-04-xx-xx-xx-002
1 Plug M 16 x 1.5 (DIN 43650/A), ½ NPT and plug DIN 43650/C
2 Plug M 12 x 1
3 Cable (rd = red, wh = white, gn = green)
For electrical connection provided by customer use only shielded cable
Plug M 16 x 1.5 (DIN 43650/A), ½ NPT
Plug DIN 43650/C
Plug M 12x1 5 m cable, only analog output
Plug-in housing Coupling nut Reference pressure line
Plug-in jack with gasket Connector with gasket
Coupling nut Operating potentiometer (inner)
Plug with O-ring
Operating potentiometer (inner)
DIN 43 650/A (M16 / ½ NPT)
DIN 43650/C
M12x1
5 m
1
1
1
2
2
3
3
4
5
+
+
+
+
PE
rd
wh
gn
3
3
2
1
1+
2
12...30 V
12...30 V
12...30 V
4...20 mA
4...20 mA
4...20 mA
1
2
3
1+
3-