EasyManua.ls Logo

Endress+Hauser Cerabar T PMC131 - Device Identification; Model and Configuration Codes

Endress+Hauser Cerabar T PMC131
24 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Cerabar T
4 Endress+Hauser
Measuring system
P01-PMx13xxx-14-xx-xx-xx-002
1 PMP131, PMP135: current output with transmitter power supply unit, e.g. RN 221N from Endress+Hauser
2 PMP131, PMP135: switch output with load, e.g. PLC, DCS, relay
3 PMC131: voltage output with transmitter power supply unit, e.g. RIA452 from Endress+Hauser
4 PMC131: current output with transmitter power supply unit, e.g. RN 221N from Endress+Hauser
Input
Measured variable Absolute pressure or gauge pressure
Measuring range up to 400 bar/6000 psi, see Page 18, "Ordering information" section
Output
Analog output (PMC131, PMP131, PMP135)
Output signal 4 to 20 mA
Load PMC131
R
Lmax
[Ω] (U
S
– 1 V) / 0.02 A
PMP131 and PMP135 (current output)
R
Lmax
[Ω] (U
S
– 12 V) / 0.02 A (R
Lmax
: Maximum load resistance, U
S
: Supply voltage)
PMP131 (voltage output)
Load resistance R
Lmax
5 kΩ, current consumption 6 mA
Switch output (PMP131, PMP135)
Output signal PNP switch output (positive voltage signal), rate depends on power supply voltage
Output current Switch status ON: I
a
500 mA
Switch status OFF: I
a
1 mA
4...20 mA
0...10 V
4...20 mA
p
pp
p
+
+
+
+
1
2
3
4
R

Table of Contents