CM14 Troubleshooting
29
Diagnostics
code
Message text Description
F100 Sensor comm.
No sensor communication.
Possible reasons:
No sensor connection
Faulty sensor connection
Short-circuit in the sensor cable
Short-circuit in the neighboring channel
Sensor firmware update canceled with an error
Remedy:
Check sensor cable connection
Check sensor cable for short-circuit
Replace sensor
Restart firmware update
Contact service
F118 Glass crack
Sensor glass breakage alarm.
Impedance of glass membrane too low.
Remedy:
Inspect glass electrode for hair-line cracks and breakage
Check the medium temperature
Check the electrode plug-in head for moisture and dry if necessary
Replace sensor
F120 Sensor ref.
Sensor reference alarm.
Impedance of reference too low.
Remedy:
Inspect glass electrode for hair-line cracks and breakage
Check the medium temperature
Check the electrode plug-in head for moisture and dry if necessary
Replace sensor
F124 Sensor glass
Alarm, sensor glass limit value exceeded.
Impedance of glass membrane too high.
Remedy:
Check pH sensor, replace if necessary
Check glass limit value and correct if necessary
Replace sensor
F142 Sensor signal
Sensor check.
No conductivity display.
Possible reasons:
Sensor in air
Sensor defective
Remedy:
Check sensor installation
Replace sensor
F143 Self test
Sensor self-test error.
Remedy:
Replace sensor
Contact service
F845 Device id Faulty hardware configuration