Commissioning CM14
18
Parameter Configuration options Description
Out main 0/4 mA Numerical value 0.000 to 99 999
1.0 pH
Physical value corresponding to the lower range
limit of the analog output.
If the set value is undershot, the current output
goes to the saturation current 0/3.8 mA.
Out main 20 mA Numerical value 0.000 to 99 999
12 pH
Physical value corresponding to the upper
range limit of the analog output.
If the set value is exceeded, the current output
goes to the saturation current 20.5 mA.
Out temp 0/4 mA Numerical value–50 to 250 °C
–5 °C
Temperature corresponding to the measuring
range lower limit of the temperature input.
If the set value is undershot, the current output
goes to the saturation current 0/3.8 mA.
Out temp 20 mA Numerical value–50 to 250 °C
100 °C
Temperature corresponding to the measuring
range upper limit of the temperature input.
If the set value is exceeded, the current output
goes to the saturation current 20.5 mA.
Damping main value 0 to 60 s
0 s
Configure the damping for the low-pass
filtering of the input signals.
Extended setup Advanced settings for the device, such as the
relay, limit values etc.
The functions are described in the following
chapter, .
Manual hold Off, On Function to hold the current and relay outputs
6.4 Extended configuration (Extended setup menu)
Press the 'E' button during operation to call up the main menu. Press the '+' button during
operation to navigate to the Setup menu. Press the 'E' key to open the menu. Navigate to the
Extended Setup menu and open the menu by pressing the 'E' key. Select the "x Back" option at
the end of each menu/submenu to navigate one level higher in the menu structure.
Parameter Configuration options Description
System General settings
Tag Free text
Max. 16 characters
Enter the device designation (tag).
Temp. Unit
°C
°F
Configures the temperature unit.
Hold release 0 to 600 s
0 s
Configuration that specifies how long a device
hold continues once the hold condition no
longer applies.
Alarm delay 0 to 600 s
0 s
Delay time for issuing an alarm. This suppresses
alarm conditions that occur for a period shorter
than the alarm delay time.
Input Settings of the device