Cerabar M, Deltabar M, Deltapilot M / 4...20 mA HART Installation
Endress+Hauser 19
Flow measurement in steam
P01-PMD55xxx-11-xx-xx-xx-001
Measuring layout for flow measurement in steam
1 Condensate traps
2 Orifice plate or pitot tube
3 Shut-off valves
4Deltabar M
5Separator
6 Drain valves
7 Three-valve manifold
Mount the Deltabar M below the measuring point.
Mount the condensate traps at the same level as the tapping points and at the same distance to
the Deltabar M.
Prior to commissioning, fill the impulse piping to the height of the condensate traps.
Flow measurement in liquids
P01-PMD55xxx-11-xx-xx-xx-002
Measuring layout for flow measurement in liquids
1 Orifice plate or pitot tube
2 Shut-off valves
3Deltabar M
4Separator
5 Drain valves
6 Three-valve manifold
Mount the Deltabar M below the measuring point so that the impulse piping is always filled with
liquid and gas bubbles can run back into the process piping.
When measuring in media with solid parts, such as dirty liquids, installing separators and drain
valves is useful for capturing and removing sediment.
+
+