EasyManua.ls Logo

Endress+Hauser Deltapilot M FMB51 - 6.8 Electrical Differential Pressure Measurement

Endress+Hauser Deltapilot M FMB51
134 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Cerabar M, Deltabar M, Deltapilot M / 4...20 mA HART Commissioning
Endress+Hauser 73
6.8 Electrical differential pressure measurement with
gauge pressure sensors (Cerabar M or Deltapilot M)
Example:
In the example given, two Cerabar M or Deltapilot M devices (each with a gauge pressure sensor)
are interconnected. The pressure difference can thus be measured using two independent Cerabar
M or Deltapilot M devices.
!
Note!
For a description of the parameters mentioned Chap. 10.2 "Description of parameters".
P01-XMX5Xxx-14-xx-xx-xx-001
1 Shut-off valves
2 e.g. filter
Fieldgate
FXA520
FXN 520
FXN 520
Multidrop-Connector
FXN520
Cerabar M/
Deltapilot M
Cerabar M/
Deltapilot M
Description
Adjustment of the Cerabar M/Deltapilot M on the high pressure side
1 Select the "Pressure" measuring mode via the "Measuring mode" parameter.
Menu path: Setup Measuring mode
2 Select a pressure unit via the "Press eng. unit" parameter, here "mbar" for example.
Menu path: Setup Press. eng. unit
3 The Cerabar M/Deltapilot M is unpressurized, perform position adjustment, see ä 56.
4 Switch on burst mode via the "Burst mode" parameter
Menu path: Expert Communication HART Config.
5 Set the output current to "Fixed" 4.0 mA via the "Current mode" parameter.
Menu path: Expert Communication HART Config
6 Configure an address 0 via the "Bus address" parameter, e.g. bus address = 1
(HART 5.0 master: Range 0 to 15, where address = 0 calls up the "Signaling" setting; HART 6.0 master: range 0 to
63)
Menu path: Expert Communication HART Config.
Description
Adjustment of the Cerabar M/Deltapilot M on the low pressure side (the differential is generated in this
device)
1 Select the "Pressure" measuring mode via the "Measuring mode" parameter.
Menu path: Setup Measuring mode
2 Select a pressure unit via the "Press eng. unit" parameter, here "mbar" for example.
Menu path: Setup Press. eng. unit