Deltapilot S XA00286P-C
11
Temperature tables
Connection data
Type of protection Temperature class Process temperature Ambient temperature
(Housing)
II 1/2 G
Ex ia IIC T6...T4 Ga/Gb
T6 80 °C –40 °C Ta +40 °C
T4 100 °C –40 °C Ta +70 °C
The process temperatures refer to the temperature at the separation membrane of FMB70
(do not exceed the max. ambient temperature at the housing).
Type of protection Ingress protection
of housing
Max. surface temperatur
under fault conditions
Ambient temperature
(Housing)
II 1/2 D
Ex ia IIIC T85 °C Da/Db
IP66/67 +85 °C –40 °C Ta +70 °C
Type of protection Electronic insert:
4-20 mA HART PROFIBUS PA, FOUNDATION Fieldbus
II 1/2 G
Ex ia IIC T6...T4 Ga/Gb
Ui 30 V DC
li 300 mA
Pi 1 W
Ci 11,8 nF
Li 225 μH
(Order code, Position 2
"Output; Operation": A, B, C)
or
Li = 0 (Order code, Position 2
"Output; Operation": D, E, F)
Ui 17.5 V DC
li 500 mA
Pi 5.5 W
or
Ui 24 V DC
li 250 mA
Pi 1.2 W
Ci 5 nF
Li 10 μH
(suitable for connection to a fieldbus system
according to the FISCO model)
Type of protection Electronic insert:
4-20 mA HART PROFIBUS PA, FOUNDATION Fieldbus
II 1/2 D
Ex ia IIIC T85°C Da/Db
Ui 30 V DC
li 300 mA
Pi 1 W
Ci 11,8 nF
Li 225 μH
(Order code, Position 2
"Output; Operation": A, B, C)
or
Li = 0 (Order code, Position 2
"Output; Operation": D, E, F)
Ui 17.5 V DC
Ii 500 mA
Pi 5.5 W
(suitable for connection to a fieldbus system
according to the FISCO model)
or
Ui 24 V DC
Ii 250 mA
Pi 1.2 W
Ci 5 nF
Li 10 μH