Parameter description (Local operation and FieldCare) Cerabar S/Deltabar S/Deltapilot S FOUNDATION Fieldbus
136 Endress+Hauser
P01-xxxxxxxx-19-xx-xx-en-080
Fig. 37: Quick Setup menu for the "Pressure" measuring mode
1
1
2
1
1) Displayvia on-site display only
2) Displayvia
digital communication
Deltabar S,
Cerabar S gauge pressure sensors,
Deltapilot S
Cerabar S absolut pressure sensors
POS. ZERO ADJUST
DAMPING VALUE
POS.INPUT VALUE
Flow
Pressure
Level
OPERATING MENU
Measured value
GROUP SELECTION
LANGUAGE MEASURING MODE
QUICK SETUP
MEASURING MODE
Table 3: QUICK SETUP "Pressure"
Parameter name Description
This menu displays the most important parameters for the "Pressure" measuring mode.
Prerequisite:
MEASURING MODE = Pressure
Note:
See also
ä 144 ff, Table 7: OPERATING MENU SETTINGS BASIC SETUP "Pressure"
ä 183, Table 15: OPERATING MENU SETTINGS EXTENDED SETUP "Pressure"
ä 200 ff, Table 27: OPERATING MENU PROCESSINFO PROCESS VALUES "Pressure"
ä 104 ff, Section 8 "Pressure measurement (via local operation and FieldCare)".
MEASURING MODE
Selection
Select the measuring mode.
The operating menu is structured according to the selected measuring mode.
!
Note!
When the measuring mode is changed, no conversion takes place. The digital output
value of the Analog Input Block OUT no longer displays the same value as the local
operation or the MEASURED VALUE. The device has to be recalibrated if the measuring
mode is changed. ä 104, Section 8 "Pressure measurement (via local operation and
FieldCare)".
Options:
Pressure
Level
•Deltabar S: Flow
Factory setting:
Cerabar S and Deltabar S: Pressure
Deltapilot S: Level