Cerabar S/Deltabar S/Deltapilot S FOUNDATION Fieldbus Troubleshooting
Endress+Hauser 209
12 Troubleshooting
In this chapter the parameter text as well as the parameter name are indicated. In FF configuration
programs only the parameter text is displayed (exeption: in the NI-FBUS configurator you can select
if the parameter text or the parameter name is displayed).
P01-xxxxxxxx-07-xx-xx-en-001
12.1 Messages
The following table lists all the messages that can occur. The device makes a distinction between
the message types "Alarm", "Warning" and "Error". For "Error"-type messages, you can enter
whether the device should react as in the event of an alarm or as in the event of a warning. See
"Message type/NA 64" column and Section 12.2 "Response of outputs to errors".
In addition, the "Message type/NA 64" column classifies the messages in accordance with NAMUR
Recommendation NA 64:
Break down: indicated with "B"
Maintenance need: indicated with "C" (check request)
Function check: indicated with "I" (in service)
Message display:
Local operation:
The measured value display shows the message with the highest priority.
See the "Priority" column.
The Alarm Status/ALARM_STATUS parameter shows all the messages present in descending
order of priority. You can scroll through all the messages pending using the S or O key.
FieldCare: The Alarm Status/ALARM_STATUS parameter shows the message with the highest
priority. See the "Priority" column.
TminALARM WINDOW
(334)
Entry
Customer-specific process monitoring – enter lower temperature limit.
You can use the SELECT ALARMTYPE parameter to enter how the device responds if the
operating pressure undershoots the specified value.
See also these Operating Instructions, Section 12.1 "Messages" Table, code E732 and
Section 12.2 "Response of outputs to errors".
Factory setting:
Lower sensor temperature application limit – 10 K
(For the lower temperature application limit, see Tmin SENSOR)
Tmax ALARM WINDOW
(335)
Entry
Customer-specific process monitoring – enter upper temperature limit.
You can use the SELECT ALARMTYPE parameter to enter how the device responds if the
operating pressure undershoots the specified value.
See also these Operating Instructions, Section 12.1 "Messages" Table, code E733 and
Section 12.2 "Response of outputs to errors".
Factory setting:
Upper sensor temperature application limit +10 K
(For the upper temperature application limit, see Tmax SENSOR)
Table 35: OPERATING MENU DIAGNOSTICS USER LIMITS
Parameter name Description
Parametertext Parametername
Linearization/LINEARIZATION