EasyManua.ls Logo

Endress+Hauser Deltapilot S FMB70 - Page 78

Endress+Hauser Deltapilot S FMB70
238 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Parameter description (FF configuration program) Cerabar S/Deltabar S/Deltapilot S FOUNDATION Fieldbus
78 Endress+Hauser
Counter: T > Tmax/
COUNTER_TMAX
Display
Index: 29
Data type: unsigned16
Access: read only
Displays the number of times the specified temperature range of the sensor has been
overshot. You can reset this counter via the Reset Peakhold/RESET_PEAK_HOLD
parameter.
Max. Meas. Temp./
MAX_MEASURED_
TEMP
Display
Index: 30
Data type: float
Access: read only
Displays the highest measured temperature in the sensor (peak hold indicator). You can reset
this maximum indicator via the Reset Peakhold/RESET_PEAK_HOLD parameter.
Counter: T < Tmin/
COUNTER_TMIN
Display
Index: 31
Data type: unsigned16
Access: read only
Displays the number of times the specified temperature range of the sensor has been
undershot. You can reset this counter via the Reset Peakhold/RESET_PEAK_HOLD
parameter.
Min. Meas. Temp./
MIN_MEASURED_
TEMP
Display
Index: 32
Data type: float
Access: read only
Displays the lowest measured temperature in the sensor (peak hold indicator). You can reset
this maximum indicator via the Reset Peakhold/RESET_PEAK_HOLD parameter.
Pcb Count: T > Tmax/
ELECTRONIC_OVER_
TEMP_COUNTER
Display
Index: 33
Data type: unsigned16
Access: read only
Displays the number of times the specified temperature range of the electronics has been
overshot.
Pcb Max. Temp./
ELECTRONIC_OVER_
TEMPERATURE
Display
Index: 34
Data type: float
Access: read only
Displays the highest electronics temperature measured.
Pcb Count: T< Tmin/
ELECTRONIC_UNDER_
TEMP_COUNTER
Display
Index: 35
Data type: unsigned16
Access: read only
Displays the number of times the specified temperature range of the electronics has been
undershot.
Pcb Min. Temp./
ELECTRONIC_UNDER_
TEMPERATURE
Display
Index: 36
Data type: float
Access: read only
Displays the lowest electronics temperature measured.
Service Transducer Block
Parameter Description

Table of Contents