Electrical connection Flowmeter Proline 200
20 Endress+Hauser
6.
L
WARNING
Housing degree of protection may be voided due to insufficient sealing of the
housing.
Screw in the screw without using any lubricant. The threads on the cover are coated
with a dry lubricant.
Firmly tighten the cable glands.
7. Reverse the removal procedure to reassemble the transmitter.
Removing a cable
213 4
3 (0.12)
A0032240
To remove a cable from the terminal, use a flat-blade screwdriver to push the slot between
the two terminal holes while simultaneously pulling the cable end out of the terminal.
5.2.2 Connecting the remote version
L
WARNING
Risk of damaging the electronic components!
Connect the sensor and transmitter to the same potential equalization.
Only connect the sensor to a transmitter with the same serial number.
The following procedure (in the action sequence given) is recommended for the remote
version:
1. Mount the sensor and transmitter.
2. Connect the connecting cable for the remote version.