Commissioning Flowmeter Proline 200
38 Endress+Hauser
Configuring the analog input (AI)
6. Configuring the analog input (AI).
Configuring the external compensation
7. With Wet Steam Detection/Measurement application package:
In the Steam quality parameter, select the Calculated value option.
For detailed information on the basic conditions for wet steam applications, see the
Special Documentation.
8.7.2 Liquid application
User-specific liquid, e. g. heat carrier oil
Select medium
Navigation:
Setup → Medium selection
1. Call up the Medium selection wizard.
2. In the Select medium parameter, select the Liquid option.
3. In the Liquid type parameter, select the User-specific liquid option.
4. In the Enthalpy type parameter, select the Heat option.
Heat option: Non-flammable liquid that serves as a heat carrier.
Calorific value option: Flammable liquid whose combustion energy is calculated.
Configuring fluid properties
Navigation:
Setup → Advanced setup → Medium properties
5. Call up the Medium properties submenu.
6. In the Reference density parameter, enter the reference density of the fluid.
7. In the Reference temperature parameter, enter the fluid temperature associated with
the reference density.
8. In the Linear expansion coefficient parameter, enter the expansion coefficient of the
fluid.
9. In the Specific heat capacity parameter, enter the heat capacity of the fluid.
10. In the Dynamic viscosity parameter, enter the viscosity of the fluid.