Mounting Levelflex FMP55 HART
24 Endress+Hauser
6.2.3 Mounting the "Sensor remote" version
This section is only valid for devices of the version "Probe Design" = "Sensor remote"
(feature 600, option MB/MC/MD).
For the version "Probe design" = "Sensor remote" the following is supplied:
The probe with the process connection
The electronics housing
The mounting bracket for wall or pipe mounting of the electronics housing
The connection cable (length as ordered). The cable has got one straight and one angled
plug (90°). Depending on the local conditions the angled plug can be connected at the probe
or at the electronics housing.
CAUTION
The plugs of the connection cable may be damaged by mechanical stress.
Mount the probe and the electronics housing tightly before connecting the cable.
Lay the cable such that it is not exposed to mechanical stress. Minimum bending radius:
100 mm (4").
When connecting the cable: Connect the straight plug before the angled one. Torque for
both coupling nuts: 6 Nm.
If the meeasuring point is exposed to strong vibrations, an additional locking compound
(e.g. Loctite 243) can be applied at the plug connectors.