Functional Safety Manual
for the Memosens transmitter Liquiline M CM42 SIL
Version: Page:
2.0
29 of 72
e vorbehalten. Das Kopieren dieses Dokuments und die Verwendung von Teilen aus diesem Dokument ist
ohne schriftliche Genehmigung der Endress+Hauser Conducta GmbH+Co. KG nicht erlaubt.
All rights reserved. Passing on and copying of this document, use and communication of its contents not
permitted without written authorization from Endress+Hauser Conducta GmbH+Co. KG.
CAL hold off
SETUP hold off
DIAG hold off
Logbooks on
Lab device off
Correct hardware and software versions See chapter 1.3
Lower Limit current output 1 and 2 Identical
Upper Limit current output 1 and 2 Identical
The fastest way to default settings required for SIL mode: Use factory settings (DIAG –
Service – Factory default).
The following screenshots give an impression how it looks like on the display:
First you probably start with no sensor connected:
Then after you have connected the sensor and the sensor has been found and the first
measurement has been done, you see: