Functional Safety Manual
for the Memosens transmitter Liquiline M CM42 SIL
Version: Page:
2.0
33 of 72
Alle Recht
e vorbehalten. Das Kopieren dieses Dokuments und die Verwendung von Teilen aus diesem Dokument ist
ohne schriftliche Genehmigung der Endress+Hauser Conducta GmbH+Co. KG nicht erlaubt.
All rights reserved. Passing on and copying of this document, use and communication of its contents not
permitted without written authorization from Endress+Hauser Conducta GmbH+Co. KG.
Now the menu looks like:
4.5 The SIL mode – Active safe state mode
This mode always displays the reason for its activation. It is given as a hexadecimal
number and looks like
In this example the reason is 00000003, which means: "user has explicitly deactivated
the SIL measurement mode (via safe adjustment)".
Here is the table of reasons for the activation of the active safe state and its meaning:
(below each reason, there is short help on what to do if this diagnostic code appears)