4.7.2 1 point temperature calibration
First enter the menu using the CAL menu in SIL measuring mode and follow the
instructions given on the display carefully.
Again: Only the safe screens must be used to document values. See the remarks given
with the 2 point pH calibration in chapter 4.7.1.
The next screen shows the raw temperature measured. Compare this to your reference
measurement and document both measured temperatures.