Functional Safety Manual
for the Memosens transmitter Liquiline M CM42 SIL
Version: Page:
2.0
45 of 72
e vorbehalten. Das Kopieren dieses Dokuments und die Verwendung von Teilen aus diesem Dokument ist
ohne schriftliche Genehmigung der Endress+Hauser Conducta GmbH+Co. KG nicht erlaubt.
All rights reserved. Passing on and copying of this document, use and communication of its contents not
permitted without written authorization from Endress+Hauser Conducta GmbH+Co. KG.
4.7.2 1 point temperature calibration
First enter the menu using the CAL menu in SIL measuring mode and follow the
instructions given on the display carefully.
Again: Only the safe screens must be used to document values. See the remarks given
with the 2 point pH calibration in chapter 4.7.1.
The next screen shows the raw temperature measured. Compare this to your reference
measurement and document both measured temperatures.