Electrical connection Liquiphant FTL62
14 Endress+Hauser
A0039520
12 Cover with securing screw
6.2.2 Connecting protective earth (PE)
The protective earth conductor at the device must only be connected if the device's operating
voltage is ≥ 35 V
DC
or ≥ 16 V
AC
eff.
When the device is used in hazardous areas, it must always be included in the potential
equalization of the system, irrespective of the operating voltage.
The plastic housing is available with or without an external protective earth connection
(PE). If the operating voltage of the electronic insert is < 35 V, the plastic housing has no
external protective earth connection.
6.3 Connecting the device
Housing thread
The thread of the electronics and connection compartment is coated with lubricant
varnish.
Avoid additional lubrication.
6.3.1 2-wire AC (electronic insert FEL61)
• Two-wire AC version
• Switches the load directly into the power supply circuit via an electronic switch; always
connect in series with a load
• Functional testing without level change
A functional test can be performed on the device using the test button on the electronic
insert.
Supply voltage
U = 19 to 253 V
AC
, 50 Hz/60 Hz
Residual voltage when switched through: typically 12 V
Observe the following as per IEC/EN61010-1: Provide a suitable circuit breaker for the
device, and limit the current to 1 A, e.g. by installing a 1 A fuse (slow-blow) in the phase
(not the neutral conductor) of the supply circuit.
Power consumption
S ≤ 2 VA