Liquiphant FTL62 Electrical connection
Endress+Hauser 13
5.2.4 Aligning the cable entry
A0042214
11 Housing with external locking screw and drip loop
The locking screw is not tightened when the device is delivered.
1. Loosen the external locking screw (maximum 1.5 turns).
2. Turn the housing, align the cable entry.
Avoid moisture in the housing, provide a loop to allow moisture to drain off.
3. Tighten the external locking screw.
6 Electrical connection
6.1 Required tool
• Screwdriver for electrical connection
• Allen key for screw of cover lock
6.2 Connecting requirements
6.2.1 Cover with securing screw
In the case of devices for use in the hazardous area with a certain type of protection, the cover
is sealed by a securing screw.
NOTICE
If the securing screw is not positioned correctly, the cover cannot provide secure sealing.
‣
Open the cover: slacken the screw of the cover lock with a maximum of 2 turns so that the
screw does not fall out. Fit the cover and check the cover seal.
‣
Close the cover: screw the cover securely onto the housing, making sure that the securing
screw is positioned correctly. There should not be any gap between the cover and housing.