Liquiphant FTL62 Operation options
Endress+Hauser 33
7.2 Elements on the electronic insert
MIN
>0,7
MAX
>0,5
COM
U = 9...20 V DC
4
3 5 8
7
6
L-L+
1
2
3
4
5
6
7
8
1
2
3
4
5
6
7
8
9
A0037705
 27 Example of electronic insert FEL64DC
1 COM interface for additional modules (LED module, Bluetooth module)
2 LED, red, for warning or alarm
3 LED, yellow, switch status
4 LED, green, operational status (device is on)
5 Test button, activates functional test
6 DIP switch for setting density 0.7 or 0.5
7 Terminals (3 to 8), relay contact
8 Terminals (1 to 2), power supply
9 DIP switch for configuring MAX/MIN safety mode
7.3 Heartbeat diagnostics and verification with Bluetooth® wireless
technology
7.3.1 Access via Bluetooth® wireless technology
1
2
A0033411
 28 Remote operation via Bluetooth® wireless technology
1 Smartphone or tablet with SmartBlue app
2 Device with optional Bluetooth module

Other manuals for Endress+Hauser Liquiphant FTL62