Mounting Liquistation CSF48
26 Endress+Hauser
Connecting the suction line on version with vacuum pump
1
2
3
4
A0013707
 14 Connecting the suction line from the side (as-
delivered state)
1 Hose
2 Fixing clip for hose gland
3 Thread adapter nut
4 Hose gland
A0013708
 15 Suction line connected from below
Modification of the suction line from side connection to connection from below
1. Unscrew the thread adapter nut (item 3).
2. Unscrew the hose gland (item 4) from the side panel.
3. Fit the hose gland in the fixing clamp (item 2) as illustrated.
4. Screw the hose tight from above.
5. Attach the hose adapter supplied to the suction line and screw it onto the hose gland
from below.
6. Insert the dummy plugs supplied.