Liquistation CSF48 Mounting
Endress+Hauser 27
Connecting the suction line on version with peristaltic pump
1
2
3
4
A0013705
 16 Connecting the suction line from the side (as-
delivered state)
1 Small thread adapter nut
2 Hose
3 Thread adapter nut
4 Hose gland
1
A0013706
 17 Suction line connected from below
Modification of the suction line from side connection to connection from below
1. Unscrew the thread adapter nut (item 3) and the hose gland (item 4) from the side
panel.
2. Unscrew the small thread adapter nut (item 1) and remove the hose.
3. Fit a hose adapter on the hose.
4. Secure the hose adapter with a worm drive hose clip.
5. Connect the suction line from below as illustrated.
6. Insert the dummy plugs supplied.