A0027502
29 Function code
To make it easier for you to select and find function
groups and functions, a code for the corresponding
field is displayed for each function → 28
The structure of this code is illustrated in → 29.
The function groups are indicated as letters in the first
column (see the names of the function groups). The
functions of the individual groups are displayed
incrementally by row and by column.
Factory settings
The first time the device is switched on, the factory setting is set for all the functions. The
table below provides an overview of the most important settings.
All other factory settings can be found in the description of the individual function groups
in the "System configuration" section (the factory setting is highlighted in bold).
Function Factory setting
Type of measurement Concentration of free chlorine/total chlorine in
mg/l
Temperature measurement in °C
pH value (EP version)
Sensor setting CCS140 for free chlorine
Alarm contact Steady contact
Alarm delay Setting in minutes
Error current for alarm 22 mA
Check functions* Off. Can be switched on if required
Limit value 1 and 2 for chlorine/chlorine dioxide 0.5 mg/l
Limit value 1 and 2 for pH* pH 7.2
Limit value 1 and 2 for ORP* 750 mV
Limit value 1 and 2 for temperature 50 °C
Current outputs 1 and 2 4 to 20 mA
Current output 1: measured value for 4 mA signal current 0.00 mg/l
Current output 1: measured value for 20 mA signal current 2.00 mg/l
Current output 2: temperature value for 4 mA signal current* 0 °C
Current output 2: temperature value for 20 mA signal current* 50 °C
* with appropriate version