Liquisys M CCM223/253 Commissioning
Endress+Hauser 77
Coding Field Range of
adjustment
(factory settings
in bold)
Display Info
R285 Enter minimum
switch-on time
t
ON
0.3 s
0.1 to 5.0 s
R285
s
Min.PTime
0.3
A0002112-EN
R286 Enter motor run
time
60 s
10 to 999 s
R286
s
MotorTime
60
A0002113-EN
Motor run time from "fully
closed" state to "fully open"
state.
R287 Enter neutral
zone
10 %
0 to 40 %
R287
%
NeutrZone
10
A0002114-EN
7.4.8 Service
Coding Field Range of
adjustment
(factory settings in
bold)
Display Info
S
SERVICE
function group
S
SERVICE
A0008408-EN
Service function settings.
S1 Select language ENG = English
GER = German
FRA = French
ITA = Italian
NL = Dutch
ESP = Spanish
S1
Language
ENG
A0008409-EN
The option selected only
applies for the fault-
signaling contact, not for the
error current.
S2 Configure a hold S+C = hold during
configuration and
calibration
Cal = hold during
calibration
Setup = hold during
configuration
None = no hold
S2
Auto HOLD
S+C
A0008413-EN
S = setup
C = calibration
S3 Manual hold Off
On
S3
Man. HOLD
off
A0008414-EN
The setting is retained even
in the event of a power
failure.