Commissioning Liquisys M CCM223/253
78 Endress+Hauser
Coding Field Range of
adjustment
(factory settings in
bold)
Display Info
S4 Enter hold dwell
period
10 s
0 to 999 s
S4
s
Cont.Time
10
A0008415-EN
S5 Enter SW
upgrade release
code (Plus
Package)
0000
0000 to 9999
S5
0000
PlusCode
A0008416-EN
The code can be found on the
nameplate.
If an incorrect code is
entered, you are taken back
to the measurement menu.
The number is edited with
the PLUS or MINUS key and
confirmed with the ENTER
key. "1" is displayed if the
code is active.
S6 Enter SW
upgrade release
code for
Chemoclean
0000
0000 to 9999
S6
CleanCode
0000
A0008417-EN
The code can be found on the
nameplate.
If an incorrect code is
entered, you are taken back
to the measurement menu.
The number is edited with
the PLUS or MINUS key and
confirmed with the ENTER
key. "1" is displayed if the
code is active.
S7 Order number is
displayed
S7
EK0005
order
A0006005
If the device is upgraded, the
order code is changed
automatically.
S8 Serial number is
displayed
S8
12345678
SerNo
A0008420-EN
S9 Reset the device
to the basic
settings
No
Sens = sensor data
Facty = factory
settings
S9
S.Default
no
A0008421-EN
Sens = last calibration is
deleted and is reset to factory
setting.
Facty = all data (apart from
A1 and S1) are deleted and
reset to the factory setting!
S10 Perform device
test
No
Displ = display test
S10
Test
no
A0008410-EN