Electrical connection Liquisys M CCM223/253
22 Endress+Hauser
G Proximity switch INS O Current input 4 to 20 mA
H Binary input 2 P Power connection
The device is approved for protection class II and is generally operated without a
protective ground connection. Circuits "E" and "C" are not galvanically isolated from
each other.
5.4 Device connection
Field device connection
Proceed as follows to connect the field device:
1. Open the housing cover to access the terminal block in the connection compartment.
2. Break through the perforation for a cable gland, mount a Pg gland and guide the
cable through this Pg gland.
3. Connect the cable in accordance with the terminal assignment.
4. Tighten the Pg gland again.
NOTICE
Non-observance could cause incorrect measurement.
‣
Make absolutely sure to protect the cable ends and terminals from moisture.
‣
Terminals marked NC may not be connected.
‣
Unmarked terminals may not be connected.
A0002276
12 Field device connection compartment sticker
Please label the sensor terminal block with the sticker provided.
Panel-mounted device connection
To connect the panel-mounted device, connect the cables in accordance with the terminal
assignment to the terminals on the rear of the device.