Commissioning Liquisys M CCM223/253
40 Endress+Hauser
7.3 Quick Setup
After power-up, you must make some settings to configure the most important functions
of the transmitter which are required for correct measurement. The following section gives
an example of this.
User entry Range of
adjustment
(factory settings in
bold)
Display
1. Press the ENTER key
2. Enter the code 22 to open access to the menus.
Press the ENTER key.
3. Press the MINUS key until you get to the "Service"
function group.
A0008408-EN
4. Press ENTER to be able to make your settings.
5. Select your language in S1, e.g. "ENG" for English.
Press ENTER to confirm your entry.
ENG = English
GER = German
FRA = French
ITA = Italian
NEL = Dutch
ESP = Spanish
A0008409-EN
6. Press the PLUS and MINUS key simultaneously to
exit the "Service" function group.
7. Press the MINUS key until you get to the "Setup 1"
function group.
A0007824-EN
8. Press ENTER to be able to make your settings for
"Setup 1".
9. In A1, select the desired sensor type.
Press ENTER to confirm your entry.
120 = CCS120
140 = CCS140
141 = CCS141
240 = CCS240
241 = CCS241
963
A0001954-EN
10. In A2, select the desired engineering unit.
Press ENTER to confirm your entry.
mg/l
ppm
ppb
A0024894-EN
11. If you have connected the INS proximity switch, you
can switch on flow monitoring of the sample
stream through the CCA250 assembly in A3.
Press ENTER to confirm your entry.
Off
INS
A0001956-EN
12. If flow briefly falls below the threshold value, you
can suppress controller switch-off by entering a
delay time in A4.
Press ENTER to confirm your entry.
0 s
0 to 2000 s