EasyManua.ls Logo

Endress+Hauser Liquisys M CCM223 - Page 41

Endress+Hauser Liquisys M CCM223
132 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Liquisys M CCM223/253 Commissioning
Endress+Hauser 41
User entry Range of
adjustment
(factory settings in
bold)
Display
13. In A5, enter the delay time for controller switch-
on.
In the case of chlorine/chlorine dioxide control, a
delay until the reception of a representative
measured value is recommended after a long
period without flow.
Press ENTER to confirm your entry.
0 s
0 to 2000 s
A5
s
On Delay
0
A0001958-EN
14. In A6, select the binary input.
Press ENTER to confirm your entry.
Hold = external
hold
Clean = cleaning
trigger
A6
Digital 1
Hold
A0001959-EN
15. In A7, enter the measured value damping.
Measured value damping causes the measured
value to be averaged over the specified number of
individual measured values (if A7 = 1, no damping
takes place).
Press ENTER to confirm your entry.
The display returns to the initial display of the
"Setup 1" function group.
1
1 to 60
A7
Damping
1
A0001960-EN
16. Press PLUS and MINUS simultaneously to switch to
the measurement mode.

Table of Contents

Related product manuals