Liquisys M CCM223/253 Commissioning
Endress+Hauser 63
A
B
A
B
A
B
A
B
A
B
1
2
3
4
5
A0025218
 38 Optimization of settings T
n
and K
p
A Actual value
B Time
1 T
n
too small
2 T
n
too large
3 K
p
too large
4 K
p
too small
5 Optimum setting
Actuating signal outputs via contacts (R247 to R2410 and R257 to R2510)
Each control contact outputs a cyclical signal whose intensity corresponds to the
controller's actuating value. A distinction is made according to the type of signal cycle:
Pulse length modulation
The bigger the calculated manipulated variable is, the longer the contact affected
remains picked up. The period T can be adjusted between 0.5 and 99 s (field R248 or
R258). Outputs with pulse length modulation are used to activate solenoid valves.
Pulse frequency modulation
The bigger the calculated manipulated variable is, the higher the switching frequency of
the contact affected. The maximum switching frequency 1/T can be set between 60 and
180 min
-1
(field R249 or R259). The on-time t
on
is constant. It depends on the set
maximum frequency and is approx. 0.5 s for 60 min
-1
and approx. 170 ms for 180 min
-1
.
170 ms. Outputs with pulse frequency modulation are used to activate directly controlled
solenoid dosing pumps.