Commissioning Liquisys M CCM223/253
84 Endress+Hauser
Coding Field Range of
adjustment
(factory
settings in
bold)
Display Info
C35 Store calibration
result?
Yes
No
New
A0002439-EN
If C34 = E xxx, then only No or
New. If New, then return to C.
If Yes/No, then return to
"Measurement".
C(4)
CALIBRATION
function group:
Zero point
CCS120
Zero point
A0006376-EN
Zero point calibration for
CCS120
C12 Zero point
calibration?
No
Yes
A0002421-EN
Only if A1 = 120
Zero point calibration:
1. Pass unchlorinated water
through the assembly.
2. Wait for 10 minutes.
3. Accept by selecting YES and
pressing ENTER
C14 Calibration status
is displayed
o. k.
E xxx
A0002423-EN
C15 Store calibration
result?
Yes
No
New
A0002424-EN
If C14 = E xxx, then only No or
New. If New, then return to C.
If Yes/No, then return to
"Measurement".