Liquisys M CCM223/253 Diagnostics and troubleshooting
Endress+Hauser 87
Error
No.
Display Tests/remedial action Alarm
contact
Error
current
Autom.
cleaning
start
PROFIBU
S
Status
Facty Facty Facty PV
1)
User User User Temp
E061 Temperature
measuring range
exceeded
Yes No No 80
44
E063 Below current output
range 1
Check measured value
and current assignment
Yes No No 80
80
E064 Current output range 1
exceeded
Yes No No 80
80
E065 Below current output
range 2
Yes No No 80
80
E066 Current output range 2
exceeded
Yes No No 80
80
E067 Set point exceeded
limit contactor 1
Check configuration Yes No No 80
80
E068 Set point exceeded
limit contactor 2
Yes No No 80
80
E069 Set point exceeded
limit contactor 3
Yes No No 80
80
E070 Set point exceeded
limit contactor 4
Yes No No 80
80
E080 Current output 1 range
too small
Increase range in "Current
outputs" menu.
Yes No X 80
X 80
E081 Current output 2 range
too small
Yes No X 80
X 80
E085 Incorrect setting for
error current
If the current range "0 to
20 mA" was selected in
field O311, the error
current may not be set to
"2.4 mA".
Yes No No 80
80
E100 Current simulation
active
Yes No X 80
X 80
E101 Service function active Switch off service function
or switch device off and
then on again.
No No X 80
X 80
E102 Manual mode active No No X 80
X 80
E106 Download active Wait for download to
finish.
No No X 80
X 80
E116 Download error Repeat download. Yes No X 0C
X 0C
E152 Measuring signal of
Cl/ClO
2
parameter
delayed or frozen
Check and service the
sensor and connection,
and replace if
necessary.
No No No 44
44