Micropilot FMR53, FMR54
28 Endress+Hauser
Supply voltage
An external power supply is necessary.
Various power supply units can be ordered from Endress+Hauser: see "Accessories" section
2-wire, 4-20mA HART, passive
"Power supply,
output"
1)
"Approval"
2)
Terminal voltage U
at device
Maximum load R, depending on the supply voltage U
0
of the power supply unit
A: 2-wire;
4-20mA HART
Non-Ex
Ex nA
Ex ic
CSA GP
10.4 to 35 V
3)
4) 5)
R [ ]?
U
0
[V]
10
10.4 21.4
20 30 35
0
500
A0017140
Ex ia / IS 10.4 to 30 V
3) 4) 5)
Ex d(ia) / XP
Ex ic(ia)
Ex nA(ia)
Ex ta / DIP
13 to 35 V
5) 6)
U
0
[V]
10
13 24
20 30 35
0
500
R [ ]Ω
A0034771
Ex ia + Ex d(ia) / IS +
XP
13 to 30 V
5) 6)
1) Feature 020 in the product structure
2) Feature 010 in the product structure
3) At ambient temperatures T
a
≤ -20 °C, a terminal voltage U ≥ 15 V is required to start the device with the minimum failure current (3.6 mA). The
start-up current can be configured. If the device is operated with a fixed current I ≥ 5.5 mA (HART Multidrop mode), a voltage U ≥ 10.4 V suffices
in the entire ambient temperature range.
4) A voltage U ≥ 12.5 V is required in the current simulation mode.
5) If the Bluetooth module is used, the minimum supply voltage increases by 3 V.
6) At ambient temperatures T
a
≤ -20 °C, a terminal voltage U ≥ 16 V is required to start the device with the minimum failure current (3.6 mA).
"Power supply,
output"
1)
"Approval"
2)
Terminal
voltage U at
device
Maximum load R, depending on the supply voltage U
0
of the power supply
unit
B: 2-wire; 4-20 mA
HART, switch
output
Non-Ex
Ex nA
Ex nA(ia)
Ex ic
Ex ic(ia)
Ex d(ia) / XP
Ex ta / DIP
CSA GP
13 to 35 V
3) 4)
U
0
[V]
10
13 24
20 30 35
0
500
R [ ]Ω
A0034771
Ex ia / IS
Ex ia + Ex d(ia) / IS + XP
13 to 30 V
3) 4)
1) Feature 020 in the product structure
2) Feature 010 in the product structure
3) At ambient temperatures T
a
≤ -30 °C, a terminal voltage U ≥ 16 V is required to start the device with the minimum failure current (3.6 mA).
4) If the Bluetooth module is used, the minimum supply voltage increases by 3 V.

Other manuals for Endress+Hauser Micropilot FMR53