Tankside Monitor NRF81 Operating menu
Endress+Hauser 213
Additional information
Read access Operator
Write access Maintenance
Safety distance
Navigation  Setup → Advanced setup → Application → Tank calculation → HTMS → Safety
distance
Description Defines the minimum level which must be present above the bottom pressure sensor
before its signal is used for the calculation.
User entry 0 to 10 000 mm
Factory setting 2 000 mm
Additional information
Read access Operator
Write access Maintenance
Hysteresis
Navigation  Setup → Advanced setup → Application → Tank calculation → HTMS → Hysteresis
Description Defines the hysteresis for the HTMS calculation. Prevents constant switching if the level is
near the switch-over point.
User entry 0 to 2 000 mm
Factory setting 50 mm
Additional information
Read access Operator
Write access Maintenance
Water density
Navigation  Setup → Advanced setup → Application → Tank calculation → HTMS → Water
density
Description Density of the water in the tank.
User entry Signed floating-point number
Factory setting 1 000 kg/m³