Loading...
Operating menu Tankside Monitor NRF81
236 Endress+Hauser
"Administration" submenu
Navigation Setup → Advanced setup → Administration
Define access code
Navigation Setup → Advanced setup → Administration → Define access code
Description Define release code for write access to parameters.
User entry 0 to 9 999
Factory setting 0
Additional information
Read access Operator
Write access Maintenance
If the factory setting is not changed or 0 is defined as the access code, the parameters
are not write-protected and the configuration data of the device can then always be
modified. The user is logged on in the Maintenance role.
The write protection affects all parameters marked with the symbol in this
document.
Once the access code has been defined, write-protected parameters can only be
modified if the access code is entered in the Enter access code parameter.
Device reset
Navigation  Setup → Advanced setup → Administration → Device reset
Description Reset the device configuration - either entirely or in part - to a defined state.
Selection Cancel
To fieldbus defaults
**
To factory defaults
Restart device
Factory setting Cancel
Additional information
Read access Operator
Write access Maintenance
** Visibility depends on communication