Tankside Monitor NRF81 Commissioning
Endress+Hauser 63
8.
A0029269
1 Datum plate
2 RTD
3 Probe position (→ 136)
Go to the Probe position parameter and enter the mounting position of the RTD
(measured from the datum plate).
This parameter, in conjunction with the measured level, determines whether the
measured temperature refers to the product or to the gas phase.
Offset for resistance and/or temperature
An offset for the resistance or the temperature can be defined in the following
submenu: Expert → Input/output → Analog IP X4-8.
• Ohms offset is added to the measured resistance before the calculation of the
temperature.
• Temperature offset after conversion is added to the measured temperature.
A0029265
1 Ohms offset
2 Temperature offset after conversion