Installation Proline Promag
26 Endress+Hauser
A
B
C
DN 2…25
( 1")1/12…
DN 40…1 05
(1 ½… ")6
A0019804
 10 Seals of process connections, Promag H 100
A Process connections with O-ring seal
B Process connections with aseptic molded seal, DN 2 to 25 (1/12 to 1")
C Process connections with aseptic molded seal, DN 40 to 150 (1 ½ to 6")
A
B
DN 2…25
( 1")1/12…
A0018782
 11 Seals of process connections, Promag H 200
A Process connections with O-ring seal
B Process connections with aseptic gasket seal
Mounting grounding rings, DN 2 to 25 (1/12 to 1")
For information on potential equalization, see the Transmitter Brief Operating
Instructions.
In the case of plastic process connections (e.g. flange connections or adhesive fittings),
additional ground rings must be used to ensure potential matching between the sensor and
the fluid. If grounding rings are not installed, this can affect the measuring accuracy or cause