Safety instructions
5
4 Safety instructions
Comply with national regulations governing mounting, electrical installation, commissioning,
maintenance and alteration procedures.
Requirements with regard to specialized technical staff for the mounting, electrical
installation, commissioning, maintenance and alteration of the measuring devices:
– trained in instrument safety
– familiar with the individual operation conditions of the devices
– for Ex-certified measuring devices: also trained in explosion protection
The measuring device is energized. Danger: Risk of electric shock! Open the measuring device
in a de-energized state only.
In the case of Ex-certified measuring devices: Only open in a de-energized state (once a delay
of 10 minutes has elapsed after switching off the power supply) or in environments which do
not have a potentially explosive atmosphere.
In the case of measuring devices in safety-related applications in accordance with IEC 61508
or IEC 61511: After alteration recommission in accordance with Operating Instructions.
Document the alteration procedure.
Use genuine parts from Endress+Hauser only.
Before removing the device: set the process in a safe condition and purge the pipe of
dangerous materials.
Hot surfaces! Risk of injury! Before commencing work, allow the system and measuring device
to cool down to a touchable temperature.
Follow the Operating Instructions for the device.
Risk of damaging electronic components!
Ensure you have a working environment protected from electrostatic discharge.
Only open housing for a brief period.
Avoid the penetration of foreign bodies, moisture or contaminants.
Replace defective seal/gaskets with genuine parts from Endress+Hauser only.
If threads are damaged or defective, the measuring device must be repaired.
Threads (e.g. of the cover for the electronics and connection compartments) must be
lubricated. Use an acid-free, non-hardening grease if an abrasion resistant dry lubricant
is non-existent.
If spacing is reduced or the dielectric strength of the measuring device cannot be guaranteed
during alteration work, perform a test on completion of the work (e.g. high-voltage test in
accordance with the manufacturer's instructions).
Service connector:
– do not connect in potentially explosive atmospheres.
– only connect to Endress+Hauser service devices.
Observe the instructions for transporting and returning the device outlined in the
Operating Instructions.
If you have any questions, contact your Endess+Hauser service organization.