EasyManuals Logo

Endress+Hauser Proservo NMS80 Functional Safety Manual

Endress+Hauser Proservo NMS80
32 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #18 background imageLoading...
Page #18 background image
Proservo NMS80/81/83
18 Endress+Hauser
L
CAUTION
Unlocking the device deactivates diagnostic functions, and the device may not be able to carry
out its safety function when unlocked.
Therefore, independent measures must be taken to ensure that there is no risk of danger while
the device is unlocked.
To unlock, proceed as follows:
1. Check the position of the hardware write protection switch (dip switch marked "WP" on the
main electronics), and set this switch to "OFF".
2. Select the sequence Setup → Advanced setup → Deactivate SIL/WHG and enter the
corresponding unlocking code (SIL: 7452) for the Reset write protection.
The "End of sequence" message indicates that the device was successfully unlocked.