Planning and
Installation
Minimum Spacing
In service cases it may be neccessary
to unplug the electronics housing from
the pipe stand. When installing, observe
the following spacing:
•
Minimum space required to unplug
the electronics housing: 120 mm;
all other sides 100 mm
•
Allowing 150 mm of slack cable will
permit the above without having to
disconnect the wiring.
The following minimum spacing require-
ments apply to the remote version:
•
Remote transmitter housing: 50 mm
•
Flow meter: 120 mm
Compact version
The housing can be rotated 340°,
permitting display viewing in all mounted
positions.
Within the housing, the display itself can
be rotated in 90° steps.
Remote version
Both the remote transmitter housing and
the meter housing can be rotated 340°.
min. 120 mm
ti031y30
Minimum spacing compact version
ti031y50
Electronics housing compact version
min. 120 mm
ti031y51
min. 50 mm
Minimum spacing remote version
ti031y52
Remote transmitter housing
7