Name E+H
Matrix
Index Read Write * Para-
meter
Data type Size
bytes
Storage
Class
TB Alert Key 69 X X M Unsigned8 1 S
TB Target Mode 70 X X M Unsigned8 1 S
TB Mode Block 71 X M DS-37 3 N/Cst
TB Alarm Summary 72 X M DS-42 8 D
TB Flowrate V0H0 73 X M float 4 D
TB Nominal Size V6H1 74 X X M float 4 S
not used 75...76
TB Flowrate Units V1H0 77 X X M Unsigned16 2 S
not used 78
TB Calibration Factor V6H2 79 X X M float 4 S
not used 80...81
TB Upper Sensor Limit 82 X O float 4 S
TB Lower Sensor Limit 83 X O float 4 S
not used 84...90
TB Vortex Frequency V0H1 91 X O float 4 D
not used 92...102
TB Application Medium V6H0 103 X X O Unsigned8 1 S
TB Sensor Temperature
Coefficient
V6H3 104 X X O float 4 S
TB Process
Temperature
V6H4 105 X X O float 4 S
TB Amplification V6H5 106 X X O Unsigned8 1 S
TB User Flowrate Unit V1H2 107 X X O float 4 S
not used 108...159
AI-Flow Block (AI) Flow
AI Block Object 160 X M DS-32 20 C
AI Static Revision 161 X M Unsigned16 2 N
AI Tag Description 162 X X M Octet String 32 S
AI Strategy 163 X X M Unsigned16 2 S
AI Alert Key 164 X X M Unsigned8 1 S
AI Target Mode 165 X X M Unsigned8 1 S
AI Mode Block 166 X M DS-37 3 N/Cst
AI Alarm Summary 167 X M DS-42 8 D
not used 168...169
AI Out 170 X M DS-33 5 D
AI PV Scale 171 X X M DS-36 11 S
AI Out Scale 172 X X M DS-36 11 S
not used 173
AI Channel 174 X X M Unsigned16 2 S
not used 175
AI PV Filter Time 176 X X M float 4 N
not used 177...178
AI Alarm Hysteresis 179 X X M float 4 S
not used 180
AI HI HI Limit 181 X X M float 4 S
not used 182
AI HI Limit 183 X X M float 4 S
not used 184
AI LO Limit 185 X X M float 4 S
not used 186
AI LO LO Limit 187 X X M float 4 S
not used 188...189
AI HI HI Alarm 190 X M DS-39 16 D
AI HI Alarm 191 X M DS-39 16 D
AI LO Alarm 192 X M DS-39 16 D
AI LO LO Alarm 193 X M DS-39 16 D
AI Simulate 194 X X M DS-50 6 N
not used 195...200
Prowirl 77 PROFIBUS-PA 5 Communication
Endress+Hauser 25