2.3 Transmitter wiring
Warning!
Make electrical connections with the power supply switched off!
When wiring up probes and sensors in explosion hazardous areas, observe the
instructions on the certificate and other appropriate regulations.
Rack wiringFig. 2.4 is a pin assignment diagram for the Silometer FMC 671 Z/676 Z.
Terminals z 30, b 14, d 14 and b 10 are connected internally
Inputs d2, d4 and z2, z4 are electrically isolated from the circuit and each
other.
The circuit zero of the unit ( ) is connected to the negative terminal of the
supply voltage.
Note!
Two indexing pins, at positions 1 and 22 in the rack connector ensure that Silometer
FMC 671 Z/676 Z transmitters only can be inserted at these points. The pins must
be inserted if the rack is not custom built by Endress+Hauser.
d b z
+ +
2
4
6
8
2
1
EC 47 Z
d2
d4
3
2
+
-
FEB 17 (P)
d2
d4
2
1
+
-
EC 37 Z
d2
d4
2
1
+
-
EC 17 Z
z2
z4
2
1
+
-
z2
z4
L
L+
20 V...30 V
z30
d32
d2
d4
+
d10
b10
+
+
d12
d14
0/4...20 mA
+
+
b12
b14
0/2...10 V
z18
d20
z20
z22
d24
z24
r
u
a
r
u
a
r
z26
d28
z28
u
a
+
+ +
14
16
32
EL 17 Z/EM 17 Z
z2
z4
EC 16 Z
2
1
ZA 67...
22
24
26
28
30
+
-
+
-
10
12
r
u a
18
20
r
u a
r
u a
L-
L+
Analogue
outputs
Relay
Power supply
Relay
Level,
volumen
Relay
Relay
BA063D11
Alarm relay
Non-Ex-area
Input
level
measurement
DC supply
Ex-area
Remote dialog
Deltapilot S
Current
Voltage
Rackbus
Capacitance/
Multicap probe
Capacitance/
Multicap probe
Multicap
Capacitance/
Multicap probe
Liquiphant or
Soliphant
Input
limit level
switching
FMC
671 Z
676 Z
Alarm relay
Fig. 2.4:
Pin assignment diagram for
Silometer FMC 671 Z/676 Z
Silometer FMC 671 Z/676 Z Chapter 2: Installation
13