10 Quick programming guide
10.1 Level measurement with separate level switching, capacitance probes
Start
Chapter 4, Section 4.1
- Constants for EC 37 Z/EC 47 Z or
EB 17 Z/EB 27 Z
- level is displayed at V0H0
Calibration
Chapter 4, Section 4.2 or 4.3
- After calibration level/volume is
displayed at V0H0
Set analogue output signal (optional)
Chapter 6
- Enter settings in the units selected
during calibration or linearization
Set relay parameters (optioanal)
Chapter 7
Lock parameter matrix (optional)
Chapter 4, Section 4.6
Reset parameters: V9H5
Enter sensor constants
- Zero frequency V3H5
- Sensitivity V3H6
Select mode V8H0
0 = Level and limit switching
Level or volume V3H0
»Empty« calibration V0H1
»Full« calibration V0H2
Select output range V0H3
0 = 0…20 mA/0…10 V,
1 = 4…20 mA/2…10 V
Set output damping V0H4
Set 0/4 mA value V0H5
Set 20 mA value V0H6
Set output at fault V0H7
0 = –10% (–2 mA/–1 V)
1 = +110% (+22 mA/11 V)
2 = last measurement
Set switch point relay 1 V1H0
Set fail-safe mode V1H1
0 = Minimum
1 = Maximum
Set switching hysteresis V1H2
Set status at fault V1H3
0 = de-energises
1 = depends on V0H7
Select sensor type V8H4
1 = capacitance probe
Adjust switch point V8H6
1 = 1 Hz, 5 pF
Calibrate V8H5
0 = probe uncovered
1 = probe covered
Press »E« to activate!!
Set switching delay V8H2
Set fail-safe mode V1H6
0 = Minimum
1 = Maximum
Set status on alarm V1H8
0 = de-energises
Lock parameter matrix V8H9
Silometer FMC 671 Z/676 Z Chapter 10: Quick programming guide