8 Other Applications
This chapter describes the configuration of Silometer FMC 671 Z/676 Z for other
applications. The applications described are:
»Dry Calibration« for continuous level measurement with Deltapilot probes
Level limit switching
Continuous level measurement with independent level limit detection
Level measurement with automatic calibration correction
8.1 »Dry calibration« for continuous level measurement with
Deltapilot probes
For Deltapilot probes it is possible to calibrate the Silometer FMC 671 Z/676 Z »dry« by
using the probe characteristics. For this alternative calibration you need:
the »zero frequency« and »sensitivity« of the sensors,
the »empty« level or offset at which the measurement should start
the maximum height
the density of the liquid.
The calibration is performed with an empty tank at atmospheric pressure.
Probe constantsThe probe constants are listed in table 9.3, page 55:
The zero frequency is marked »f
o
« in Hz. This is the standard frequency of the
probe at atmospheric pressure.
- Provided the sensor is unpressurized, the actual frequency, which takes into
account orientation effects, can be read at V0H8.
The sensitivity is marked »f« in Hz/mbar. This is the increase in frequency for
every mbar increase in hydrostatic pressure.
These parameters are entered at V3H5 and V3H6 respectively.
BA063E39
Max. level, e.g. 10 m,
= 980.7 mbar
Sensitivity V3H2
Pmbar at max. level
Display at max. level
Offset V3H1
P =
ρ
x g x
h mbar
Max. display e.g. 1526
"Empty"
h
Fig. 8.1:
Parameters required for
»dry calibration« of the
Silometer FMC 671 Z/676 Z,
.....
Silometer FMC 671 Z/676 Z Chapter 8: Other Applications
41