No.
V2H2
Volume
V2H3
Level
V2H4
No.
V2H2
Volume
V2H3
Level
V2H4
116
217
318
419
520
621
722
823
924
10 25
11 26
12 27
13 28
14 29
15 30
Step Matrix Entry Significance
1 V9H5 e.g. 670 Reset transmitter, see Section 4.1 (default = operating mode 1)
- Enter probe constants at V3H5 and V3H6
2 V0H1/V0H2 - Calibrate as described in Section 4.2
3 V2H1 0 Select manual entry of values
4 - »E« Register entry
5V2H21
30 Enter table entry number
6 - »E« Register entry
7 V2H3 e.g. 0 Enter volume in required units
8 - »E« Register entry
9 V2H4 00.00 Enter level for support point number
10 - »E« Register entry
11 V2H5 2
30 Second support point number will be displayed
12 - »E« Confirm input, the system jumps to V2H3,
the support point number is addet automatically
13 V2H3 Repeat steps 7 to 12 until the filling volume and level
have been entered for all support points
14 V2H0 3 Select manual linearization table
15 - »E« Press »E« to activate linearization
16 V0H0/V0H9 - Display of volume and level before linearisation
Note!
Set analogue output and relay switch point in same units as used for linearization
Manual linearization
with tabular values
Silometer FMC 671 Z/676 Z Chapter 5: Linearization
30