Calibration for
capacitance probes
Step Matrix Entry Significance
1 V9H5 671 Reset transmitter
2 - »E« Register entry
3 V8H0 2 Select level switching mode
- If 0 is entered, both level measurement (channel 1)
and level switching (channel 2) is possible
4 - »E« Register entry
5 V8H4 1 Set sensor type to capacitance probe
6 - »E« Register entry
7 V8H6 e.g. 1 Set switching point in Hz (build-up compensation)
-1 Hz
5 pF, see Table 8.1
8 - »E« Register entry
9V8H5
Either
0 Calibrate with exposed probe
10 - »E« Register entry (frequency stored at V1H5)
9V8H5
or
1
Calibrate with covered probe
10 - »E« Register entry (frequency stored at V1H5)
11 V1H4 2 If appropriate, assign relay 1 to channel 2
- Enter same values in V1H0/V1H2 as are to be found at
V1H5/V1H7
12 - »E« Register entry
13 V1H1/V1H6 e.g. 0 Select fail-safe mode
- 1 = maximum (default), 0 = minimum
14 - »E« Register entry
15 V8H2 e.g. 2 Set relay switching delay in seconds (default = 1 s)
16 - »E« Register entry
Note!
At step 9, always press button »E« even if the correct digit has already appeared in
the display.
The value in field V1H7 must be less than that in V8H6 otherwise the relay will not
switch correctly
Product Dielectric
constant
ε
r
Conductivity Switching point
with ground tube without ground tube
Solvents, oil, fuel <3 low ca. 150 mm ca. 500 mm
dry bulk solids <3 low ca. 350 mm
(cable probe)
moist bulk solids >3 medium ca. 150 mm
(cable probe)
aqueous liquids >3 high ca. 30 mm ca. 30 mm
Table 8.1:
Switching point adjustment for
default value of 1 Hz = 5 pF as a
function of product
Silometer FMC 671 Z/676 Z Chapter 8: Other Applications