Fault messages and
warnings
The fault sequence is priority-based, i.e. if a low-priority fault is present and another fault
with higher priority occurs, the fault with the highest priority is displayed. The previous
error is shown in V9H1.
Code Type Cause and Remedy
E 101-106 Fault Fault in instrument electronics
- Call Endress+Hauser Service
E 107 Fault Battery voltage too low
- Make back-up of entered parameters immediately
- Have battery changed at once by trained personel
E 201-202 Fault Fault in probe on channel 1 (f < 35 Hz; f > 3000 Hz)
- Check probe and electronic insert
E 301-302 Fault Fault in probe on channel 2 (f < 35 Hz; f > 3000 Hz)
- Check probe and electronic insert
E 400 Fault Fault in probe on channels 1 + 2
- Check probe, electronic insert and wiring
- Incorrect operating mode selected (only one probe),
operating mode V8H0
E 401 Fault Fault in probe or wiring, channel 1
- Check probe, electronic insert and wiring
- Incorrect operating mode
E 402 Fault Fault in probe or wiring, channel 2
- Check probe, electronic insert and wiring
- Incorrect operating mode
E 600 Warning PFM transmission internal code check
- can be ignored if it appears only briefly
E 601 Warning PFM transmission internal code check
- can be ignored if it appears only briefly
E 602 Warning Linearization does not rise monotonously (volume does not increase
with level)
- Check and re-enter correct values
E 604 Warning Linearization has less than two sets of values
- Enter more values
E 606 Warning No stored factory linearization (V2H6 = 0)
- Select another linearisation function. The diagnostic code can be
eliminated by pressing the »E« button in field V2H0
E 608 Warning Value in V0H5 greater than that in V0H6
- Check input
E 610 Warning Calibration fault, channel 1 (»empty« level > »full« level)
- Repeat calibration
E 613 Warning Instrument in simulation mode, channel 1
- Switch back when finished
E 614 Warning Instrument in simulation mode, channel 2
- Switch back when finished
Table 9.1:
Error messages
Silometer FMC 671 Z/676 Z Chapter 9: Trouble-Shooting
51