Self-Test issues cont.
b. There is debris or condensation trapped inside the unit, perform a FLUSH of
the unit by following the instructions in the UNDERSTANDING THE FLUSH
MODE section of this manual. Then Press the WEIGHT DOWN button to
retry.
c. Call EEC’s assistance hot line. 1-800-445-8581.
2. If upon Self-Test you get the following display:
This may indicate a number of problems. In order to test all internal lines and valves the
ADS 2000 attempts to pressurize and then checks for leaks. If it cannot pressurize then
it will give the above readout. The causes for the error are as follows:
a. No oxygen or very low oxygen pressure. Confirm there is 50 PSI of oxygen in
the Green oxygen hose that runs to the back of the ADS 2000. Press the
WEIGHT DOWN button to Retry.
b. The "To Vaporizer" and / or "From Vaporizer" connectors are loose, check
all vaporizer hoses and connections, confirm they are secure and have no
leaks. Press the WEIGHT DOWN button to Retry.
c. The vaporizer has an internal leak. To diagnose this problem connect the "To
Vaporizer" hose directly to the "From Vaporizer" hose, and Press the WEIGHT
DOWN button to Retry. If that works, it indicates that the vaporizer is the
problem.
d. You are not placing you thumb over the gas-sampling elbow that is
connected to the end of the breathing circuit during start-up. Place your thumb
over the end of the gas-sampling elbow and press the WEIGHT DOWN button
to Retry.
e. The Gas Sampling Hose is disconnected or not installed properly.
f. Low Battery - recharge unit / replace battery.
g. If the problem persists call EEC's assistance hot line. 1-800-445-8581
30 Engler Engineering Corporation ADS 2000