Self-Test cont.
3. If upon Self-Test you get the following display:
This indicates that there is leak somewhere in the system.
a. There is debris or condensation trapped inside the unit, perform a FLUSH of
the unit by following the instructions in the UNDERSTANDING THE FLUSH
MODE section of this manual. Then Press the WEIGHT DOWN button to Retry.
b. One of the hoses may be loose, check all connections and Press the WEIGHT
DOWN button to Retry.
c. The Breathing Circuit has a leak or cuff is loose, check the Breathing Circuit
for leaks, confirm it is securely connected to the Breathing Circuit Ports on the
front of the unit. Press the WEIGHT DOWN button to Retry.
d. Vaporizer has a leak or hose(s) connected to vaporizer have come loose.
e. Pop Off Valve adjustment needed. This must be completed by EEC as
specialized tools and test equipment is required. Call 800-445-8581
4. If upon Self-Test you get the following display:
This display indicates that there is a minor leak or the Safety Pop Off is
releasing at too low a pressure.
ADS 2000 Engler Engineering Corporation 31