Self-Test cont.
a. There is a loose connection to the ADS 2000 or to the vaporizer, check and
secure all connections. Press the WEIGHT DOWN button to Retry.
b. There is debris or condensation trapped inside the unit, perform a FLUSH of
the unit by following the instructions in the UNDERSTANDING THE FLUSH
MODE section of this manual. Then Press the WEIGHT DOWN button to Retry.
c. If the problem persists call EEC's assistance hot line. 1-800-445-8581
5. If upon Self-Test you get the following display:
This display indicates that there is an obstruction to the Exhale Valve or that the
Exhale Valve did not open. The following can cause this error.
a. There is an obstruction in the scavenging system, check to ensure
that a free flow of exhaust gas can pass through the scavenging
system. Then Press WEIGHT DOWN button to Retry.
b. An active scavenging system is being used and the active scavenger
valve is CLOSED or SHUT. Open the active scavenger valve and then
Press WEIGHT DOWN button to Retry.
c. Battery is low – recharge unit / replace battery.
d. If the problem persists call EEC's assistance hot line.
1-800-445-8581